Calibration
96
W-4510 Weigh Scale Technical Reference Manual
Step 2: Cal Hi on absorber
181 CAL HI, ABSORBER
CONFIRM
YES
NO (ABORT)
Figure 44: Cal on absorber screen
Procedure 31: Cal Hi on absorber—Channel #1
1. From
screen
181 CAL Hi
screen, select
YES
and press ENTER to confirm
2. From
screen
170 CAL TIME LEFT
screen, press ENTER after timeout
3. On
screen
171 EVAL DATA COLL.
, view the absorber value and measured value.
Press ENTER on the YES field to adjust the cal or press SELECT and press
ENTER to choose NO
4.
Connect the write protect jumper, JP17 and save to EEPROM
5. On
172 SAVE CALIBRATE RESULT IN EEPROM
screen, press SELECT to
choose
YES
or
NO
and press ENTER
6.
Disconnect the write protect jumper, JP17.