Diagnostics and repair
W-4510 Weigh Scale Technical Reference Manual
123
LED summary table
Table 21: Power supply board LED summary table
LED
Description
Normal condition Error condition
Recommendation
+6V
+6V DC voltage level
to electronics
ON
OFF—Electronics are not
rec6VDC voltage
required for functioning
6V on test points.
Check fuse on Power Supply
board. Check power input
terminals 1, 2.
+24V Not
used
Relay Not
used
Table 22: CPU board LED summary table
LED
Description
Normal condition Error condition
Recommendation
Mem Not
used
N/A
N/A
HART Not
used
ON
N/A
CPU Central
processing
unit on CPU board
“heartbeat”
Blinks at rate of
1time/sec.
LED does not blink. CPU
not functioning.
Check power input. Replace
CPU board.
Aux Auxiliary
input
frequency signal
indicator
Blinks if auxiliary
input present. OFF—
if no auxiliary input
present.
None
Check auxiliary input wiring
terminals 11 & 12 with a meter
for frequency signal. Check
auxiliary input equipment.
HV Not
used
N/A
N/A
Field Radiation
field
indicator
Cycles in proportion
to radiation field
intensity at detector.
ON for 10sec. for
each mR/hr, then off
for two seconds.
(Can use LED 5 that
blinks 1time/sec to
time LED 9 for field
indicator.)
None
A 1mR/hr (2,580nC/kg/hr) field
is usually required for a
measurement. Check for closed
source shutter, buildup, and
insulation.
Summary of Contents for W-4510
Page 9: ...Preface viii W 4510 Weigh Scale Technical Reference Manual Notes...
Page 15: ...Preface xiv W 4510 Weigh Scale Technical Reference Manual Notes...
Page 17: ...Preface xvi W 4510 Weigh Scale Technical Reference Manual Notes...
Page 29: ...Introduction 12 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 125: ...Calibration 108 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 145: ...Diagnostics and repair 128 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 155: ...Appendix I Parameter blocks 138 W 4510 Weigh Scale Technical Reference Manual Notes...