Setup
78
W-4510 Weigh Scale Technical Reference Manual
Procedure 19: Setup totalizer #1
1. From
004 TOTALIZER MENU
, select
TOTALIZER #1
2. From
030 TOTALIZER #1
screen, select one of the following:
•
RESET
•
PRESET
•
HOLD
•
REST/HOLD
And press ENTER
3.
If you chose
PRESET
,
From the
115 TOTALIZER #1 ENTER PRESET VALUE
screen, enter the new
value of totalizer #1 and press ENTER
4.
Connect the write protect jumper JP17
5.
Save to EEPROM
6.
Disconnect the write protect jumper, JP17. This protects the information in the
EEPROM.
Procedure 20: Setup totalizer #2
1. From
004 TOTALIZER MENU
screen, select
TOTALIZER #2
2. From
031 TOTALIZER #2
screen, select one of the following:
•
RESET
•
PRESET
•
HOLD
•
REST/HOLD
And press ENTER
3.
If you chose
PRESET
,
From the
116 TOTALIZER #2 ENTER PRESET VALUE
screen, enter the new
value of totalizer #2 and press ENTER
4.
Connect the write protect jumper JP17
5.
Save to EEPROM
6.
Disconnect the write protect jumper, JP17. This protects the information in the
EEPROM.
Summary of Contents for W-4510
Page 9: ...Preface viii W 4510 Weigh Scale Technical Reference Manual Notes...
Page 15: ...Preface xiv W 4510 Weigh Scale Technical Reference Manual Notes...
Page 17: ...Preface xvi W 4510 Weigh Scale Technical Reference Manual Notes...
Page 29: ...Introduction 12 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 125: ...Calibration 108 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 145: ...Diagnostics and repair 128 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 155: ...Appendix I Parameter blocks 138 W 4510 Weigh Scale Technical Reference Manual Notes...