
Calibration
98
W-4510 Weigh Scale Technical Reference Manual
Step 4: Absorber value setup
159 ENTER ABSORBER
VALUES.
LOW ABSORBER VALUE
HI ABSORBER VALUE
Figure 46: Absorber value setup screen
Procedure 33: Absorber value setup—Channel #1
1. From
screen
159 ENTER ABSORBER VALUES
, select
HI ABSORBER VALUE
and press ENTER
2. From
screen
148 HIGH ABSORBER screen
, enter the new value of the absorber
on high and press ENTER and then press ENTER again to confirm
3.
Connect the write protect jumper, JP17 and save to EEPROM
4.
Disconnect the write protect jumper, JP17.
Step 5: Saving the calibration information to EEPROM
After completion of step 4, all of the calibration information is in the RAM memory of the
Smart Pro. You must copy this data to the non-volatile EEPROM for long-term storage. On all
system re-boots, the EEPROM data copies into the RAM memory and becomes the current
configuration. Refer to page 46 for information concerning the upload and download
configuration program. Follow the procedure on page 93 to save the calibration information to
the EEPROM.
Summary of Contents for W-4510
Page 9: ...Preface viii W 4510 Weigh Scale Technical Reference Manual Notes...
Page 15: ...Preface xiv W 4510 Weigh Scale Technical Reference Manual Notes...
Page 17: ...Preface xvi W 4510 Weigh Scale Technical Reference Manual Notes...
Page 29: ...Introduction 12 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 125: ...Calibration 108 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 145: ...Diagnostics and repair 128 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 155: ...Appendix I Parameter blocks 138 W 4510 Weigh Scale Technical Reference Manual Notes...