Diagnostics and repair
122
W-4510 Weigh Scale Technical Reference Manual
Table 18: Power supply board test points and labels
Power Supply Test point label Description
H1 Not
used
H2 Not
used
TP1 Isolated
ground
TP2
Loop current test point 200mV/mA loop current. Referenced to isolated
ground.
Table 19: CPU board test points and labels
CPU Test point label
Description
Count
Raw input signal coming from preamp.
GND Logic
ground
U5 pin8
+5V power supply test points referenced to Logic ground.
Jumpers
The W-4510 Weigh Scale uses jumpers J1–J4 on the CPU board as division values for the
output frequency to the Smart Pro.
Note: Do not change the jumper from the current setting without consulting Ohmart Field
Service
Table 20: Jumper division values
J1
Divide by 20
J2
Divide by 10
J3
Divide by 5
J4
Divide by 2
None
Divide by 1
If the W-4510 Weigh Scale does not have a jumper, the division value is one, and the output
frequency does not change.
LED indicators
Check the basic functioning of the Ohmart electronics at the instrument with LED indicators
on the CPU board. They are visible when you remove the explosion proof housing cap.
See the tables on page 123 for a summary of the LED indications.
FIELD
Off
On
Blinking
MEM
HA
R
T
CPU
AUX
HV
FIELD
HV
AUX
CPU
HA
R
T
MEM
Figure 63: GEN2000 LED indicators
Summary of Contents for W-4510
Page 9: ...Preface viii W 4510 Weigh Scale Technical Reference Manual Notes...
Page 15: ...Preface xiv W 4510 Weigh Scale Technical Reference Manual Notes...
Page 17: ...Preface xvi W 4510 Weigh Scale Technical Reference Manual Notes...
Page 29: ...Introduction 12 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 125: ...Calibration 108 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 145: ...Diagnostics and repair 128 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 155: ...Appendix I Parameter blocks 138 W 4510 Weigh Scale Technical Reference Manual Notes...