Setup
44
W-4510 Weigh Scale Technical Reference Manual
Upper Ram transfers
The upper RAM can hold a copy of the active gauge configuration. This is useful when
making a copy of the non-volatile EEPROM. Use this screen to make transfers to and from
upper RAM. Enter a one to initiate the transfer. The transfer completes almost
instantaneously.
Procedure 2: Transfer to upper RAM
1. From
310 MEMORY BACKUP
screen, select
UPPER RAM TRANSFERS
and
press ENTER
2. From
314 UPPER RAM XFERS
screen, select
RAM TO UPPER RAM
and press
ENTER
3. In
316 RAM TO UPPER RAM TRANSFER
screen, press 1 and press ENTER to
activate the transfer
Procedure 3: Transfer to RAM
1. From
310 MEMORY BACKUP
screen, select
UPPER RAM TRANSFERS
and
press ENTER
2. From
314 UPPER RAM XFERS
screen, press SELECT to choose
UPPER RAM
TO RAM
and press ENTER
3. In
318 UPPER RAM TO RAM TRANSFER
screen, press 1 and press ENTER to
activate the transfer