Setup
W-4510 Weigh Scale Technical Reference Manual
55
Procedure 10:Setup gauge span—Channel #1
1. From
098 SET GAGE SPAN
screen, select
SET MINIMUM READING
and press
ENTER
2. From
096 MIN SPAN SETTING
screen press HELP
Screen
799
displays “CHANGING SPAN LIMITS REQUIRES GAGE
RECALIBRATION. PRESS HELP.” Press HELP to return to screen
096
and press
ENTER
3. From
096 MIN SPAN SETTING
screen, press select and type in the new minimum
span setting and press ENTER
4. Press
PREVIOUS SCREEN
to return to screen 098
5. From
098 SET GAGE SPAN
screen, select
SET MAXIMUM READING
and press
ENTER
6. From
097 MAX SPAN SETTING
screen, press select to get to the
NEW SETTING
area and type in the new max span setting and press ENTER
7.
Connect the write protect jumper JP17
8.
Save to EEPROM
9.
Disconnect the write protect jumper, JP17. This protects the information in the
EEPROM.