Setup
68
W-4510 Weigh Scale Technical Reference Manual
Get linearizer data points
The linearizer curve corrects for the inherent non-linear response of a nuclear gauge. The
value that you enter in the linearizer curve automatically compensates for the non-linear
response and then provides an accurate measurement. Since the linearizer curve stores as
one of the many parameters in the product code, it automatically applies to the calibration
and measurement process through the re-cal of each product code.
Before starting this procedure:
5
Have the process full to maximum
5
Prepare to write the datapoints on the Linearizer curve chart on page 70
5
Prepare to take process off the conveyor in increments of 10%