Calibration
W-4510 Weigh Scale Technical Reference Manual
95
Step 1: Simple cal on low
141 START SIMPLE
CAL ON EMPTY
CONFIRM
YES
NO (ABORT)
Figure 43: Simple re-cal on low screen
Procedure 30: Simple re-cal Channel #1
1. From
screen
149 CAL CHANNEL #1
, select
SIMPLE RE-CAL LOW
and press
ENTER
2. From
screen
141 START SIMPLE CAL ON EMPTY
, select
YES
and press
ENTER to confirm
3. On
screen
173 CAL TIME LEFT
, press ENTER when the counter reaches 0
4. From
screen
174 EVAL DATA COLLECT
, verify that the absorber and process
value are correct. If not correct, press SELECT for
NO
to adjust the CAL or
YES
to
continue. Press ENTER
5.
If you chose YES, screen
175
displays
6.
Connect the write protect jumper, JP17 and save to EEPROM
7. On
175 SAVE CALIBRATE RESULT IN EEPROM
screen, press SELECT to
choose
YES
or
NO
and press ENTER
8.
Disconnect the write protect jumper, JP17. This protects the information in the
EEPROM.
Summary of Contents for W-4510
Page 9: ...Preface viii W 4510 Weigh Scale Technical Reference Manual Notes...
Page 15: ...Preface xiv W 4510 Weigh Scale Technical Reference Manual Notes...
Page 17: ...Preface xvi W 4510 Weigh Scale Technical Reference Manual Notes...
Page 29: ...Introduction 12 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 125: ...Calibration 108 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 145: ...Diagnostics and repair 128 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 155: ...Appendix I Parameter blocks 138 W 4510 Weigh Scale Technical Reference Manual Notes...