Appendix I: Parameter blocks
136
W-4510 Weigh Scale Technical Reference Manual
Table 31: Screen 527 Product code block
Item# Parameter
Name
Parameter description
0
TIME CONSTANT
When the product code table entry #18 is set to zero, time constant is the time (in
XXXXX seconds) that 63% of a step change applied at the input of the digital
filter will appear out the output of the filter
When the product code table entry #18 is set to one, time constant defines the
number of readings to use to generate a measurement average. (100 readings
max)
1
FAST RESPONSE
FILTER ENTRY
(Chrontrol)
When the difference between the current gauge reading and the previous gauge
reading exceeds the value of this entry, temporarily bypasses the filter
2
SPAN
Application of gain to get normalized counts after subtraction of cal low
reference. Gain=entry/1000
3
LINEARIZER
NUMBER
Which linearizer curve to use (0–9). Normally, the curve number is the same as
the product code number. For example, Product Code #0 usually uses Curve #0,
Product Code #1 uses Curve #1, etc.
4
TEMPERATURE
COEFFICIENT
When temperature compensation enables. Entry=0.0XXXXXSGU per
degrees C or F
5
DENSITY TYPE
Enter 0=channel is not density or is straight density in SGU units. Enter
1=channel is a slurry gauge and converts to %Solids and Dry Solids. Enter
2=solutions and gives %Solids conversion
6
SOLIDS DENSITY
For density types 1 and 2 only
7
LIQUID DENSITY
For density types 1 and 2 only
8
STANDARDIZE
ABSORBER
VALUE
In units of product, frequency input block location 12
9
CALIBRATE HIGH
ABSORBER
VALUE
In units of product, frequency input block location 12
10
FREQUENCY
OUTPUT LOW
LIMIT
Value of data form the data pointer that gives min. analog output
11
FREQUENCY
OUTPUT HIGH
LIMIT
Value of data form the data pointer that gives max. analog output
12
LOW ALARM
LIMIT
Setpoint for triggering a low alert
13
HIGH ALARM
LIMIT
Setpoint for triggering a high alert
14
LOW-LOW ALARM
LIMIT
Setpoint for triggering a low alarm
15
HIGH-HIGH
ALARM LIMIT
Setpoint for triggering a high alarm
16
NUMBER OF
TABLES
Not currently used
17
COEFFICIENT
TYPE
0=Normal (Negative) Temp Coef, 1=Positive Temp Coef
18
FILTER TYPE
0=RC Simulation Filter, 1=Averaging Filter
19
SPARE
20
SPARE
21
SPARE
22
SPARE
23
SPARE
24
SPARE
Summary of Contents for W-4510
Page 9: ...Preface viii W 4510 Weigh Scale Technical Reference Manual Notes...
Page 15: ...Preface xiv W 4510 Weigh Scale Technical Reference Manual Notes...
Page 17: ...Preface xvi W 4510 Weigh Scale Technical Reference Manual Notes...
Page 29: ...Introduction 12 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 125: ...Calibration 108 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 145: ...Diagnostics and repair 128 W 4510 Weigh Scale Technical Reference Manual Notes...
Page 155: ...Appendix I Parameter blocks 138 W 4510 Weigh Scale Technical Reference Manual Notes...