Parameter Setting
page 83
rev. 00 / 04.09.2017
28.3.2 Analog Outputs
The transmitter is equipped with one analog output 0-10 V.
Default setting:
output not active
The following different functions can be assigned to the analog output.
Fig. 28-9
Analog output activation
•
Flow
The application flow rate (calculated from average flow velocity and wetted cross
section) is available on the selected analog output.
Selection/Input Options:
Output range:
>0-5 V< or >0-10 V<
Value at 0 V:
manual input
Value at 10 V:
manual input
Value at error:
>0 V< or >Hold value< or >10 V<
•
Flow Velocity
The calculated average flow velocity (calculated even from two or three sensors)
used to compute the current flow rate is available on the selected analog output.
Selection/Input Options:
Output range:
>0-5 V< or >0-10 V<
Value at 0 V:
manual input
Value at 10 V:
manual input
Value at error:
>0 V< or >Hold value< or >10 V<
•
Water Temperature
The medium temperature calculated based on the transit times is available on the
selected analog output.
Selection/Input Options:
Output range:
>0-5 V< or >0-10 V<
Value at 0 V:
manual input
Value at 10 V:
manual input
Value at error:
>0 V< or >Hold value< or >10 V<