Parameter Setting
page 103
rev. 00 / 04.09.2017
32 Parameter Menu Display
Use the display menu to determine main screen attributes.
Variable settings:
• Names of the five main screen output fields
• Decimal places of individual values
Fig. 32-1
Main screen and output fields
Output Fields
The five main screen output fields (Flow, Level, Velocity, Temperature and Total) can be speci-
fied freely in terms of name and the number of decimal places.
Note
The ASSIGNMENT of values and fields CANNOT be changed.
Example: the "Flow" field will ALWAYS indicate output values even if the name should be
changed to "Temperature".
Â
Ü
Procedure to change the name:
1. Drop down the desired output field.
2. Uncheck >Standard Name<.
Specify the new name. The name can be selected freely, however more than 16
characters will not be accepted.
3. The new name does not change the field value in the main screen.
4. Go "Back" several times to save the parameters.
F
c
To save see chapter „25.1 Save Parameters“.
Â
Ü
Procedure to change the number of decimal places:
1. Drop down the desired output field.
2. Uncheck >Default Digits<.