Start-Up Examples
page 59
Rev. 00 / 04.09.2017
2. Confirm modified parameters and path rearrangement. The display shows “Initial-
ised!” after confirmation with >Yes<.
Â
Ü
Sensor Selection
and
Specification of Mounting Angle
procedures:
1. Select menu >v-Paths<.
2. Choose the sensor type used.
3. Enter the mounting angle (+45° or -45°) and confirm.
Fig. 23-4
Enter the mounting angle (Clamp-On)
The >Distance along< field on the display indicates the distance between both sensors.
The indicated distance is always the clearance between both sensors.
All other parameters are read-only or remain to be set to default.
Â
Ü
Once entered, all required measurement place Parameters need to be saved:
1.
Use “Back” to exit the menus until >Save Parameters?< is shown on the display.
2.
Confirm >YES<. The message “Successful!” appears.
NivuFlow Mobile goes to the main screen and continues operation using the newly saved
parameters.
23.2.2 Extended Parameter Setting
More specifications:
• Pipe with inner lining
• Sedimentation within the pipe
For pipes with inner lining:
Â
Ü
Procedure:
1.
Execute steps 1 (“Menu” field, page 56) up to and including 10 (“Wall Material”;
page 58) as described in chapter „23.2.1 Simple Parameter Setting“.
2. Select the >Lining< material.
The thickness of the lining is now included in the graphics area. The NivuFlow
Mobile requires this extra value for calculations
including
the lining.