Instruction manual
NivuFlow Mobile 600
page 60
Rev. 00 / 04.09.2017
3. The lining thickness is indicated in the graphics area.
4. If moreover there should be sedimentation within the pipe enter the according
value in >Sludge Level< and confirm.
The transmitter then will subtract this sludge level from the wetted hydraulic total
area while calculating the flow rate.
5.
To save your entries go “Back” and confirm the path modification/arrangement with
>Yes<.
6. Menu >v-Paths<: choose sensor and specify the mounting angle.
Here you can read the >Distance along< and the >Path Length< values required to
install the clamp-on sensors.
24 Measurement with wet Sensors
The setting of parameters for a measurement place using wet sensors is essentially in line with
the parameter setting procedures used for clamp-on sensors.
When selecting the >Transit Time Mode<, consider that the >Path Arrangement< option
provides more variations and >Distance across< is indicated additionally when viewing the
mounting values.