Start-Up Examples
page 57
Rev. 00 / 04.09.2017
4.
Specify measurement place name and confirm with “Enter”.
5. Select transit time mode >Clamp-On<.
6.
Specify path arrangement (“diametrical \”) and number of paths (1 path).
Fig. 23-1
Select path arrangement
Hints on the Medium
If you cannot find your medium to measure within the list select “User defined”.
Another menu opens up which can be used to specify e. g. the speed of sound within the
medium.
Tipp:
Various speeds of sound can be found in the Internet or contact the manufacturer.
7. Use the selection menu to specify the medium to measure and to select/specify the
current medium temperature.
8.
Set the channel profile to “Pipe”.
The graphics area indicates a pipe featuring four input fields.
9. Enter the pipe data (example: DN1000). Two specifications are sufficient to enter
the pipe dimensions such as inside diameter and wall thickness in this example.