Installation and Connection
page 35
rev. 00 / 04.09.2017
Fig. 16-4
s
ender/re
c
e
iv
er
s
ender/re
c
e
iv
er
Example “Diametrical W” mode
16.3 Installation of wet Sensors
Important Note
Wet sensors shall be installed only by a pipeline company or a plumber. The tightness of
pipes must be guaranteed at any time.
The measurement is not influenced by pipe material or pipe wall thickness if wet sensors are
used.
The pipe sensors are designed exclusively for full pipelines. These sensors are installed in
pipelines from the outside by using a welding nozzle (90°) and a fastening element. The meas-
urement paths are always directed through the pipe centre (diametrical).
The sensors or the nozzles required to insert the sensors can be pre-installed following the
specifications. To do so, the number of paths as well as the path arrangement need to be
determined first.
The measurement place parameters (selection of measurement method and number of paths)
can be set in the >Measure Place< menu. The >v-Paths< menu is used to choose the sensor
position for each path separately. To switch to the next path choose the according tab (1 or 2)
on the top right.
The following parameters need to be specified:
• Pipe material
• Pipe outside diameter, internal pipe diameter or pipe circumference
• Pipe wall thickness
• Medium to measure
Fig. 16-5
Entering pipe dimensions and mounting angle