Instruction manual
NivuFlow Mobile 600
page 58
Rev. 00 / 04.09.2017
Fig. 23-2
Specifying pipe dimensions
As soon as inner diameter and wall thickness are specified the transmitter auto-
matically adds outside diameter and pipe circumference. The same applies after
circumference and wall thickness have been specified. The transmitter will add the
rest of the parameters.
10. Select the wall material from the menu (stainless steel).
More specifications are not required – the following parameters (lining, sludge level, ...) remain
to default state.
Once a relevant parameter in menus >Measure Place< or >v-Paths< has been changed it is
necessary to re-initialise the path arrangements in order to recalculate path lengths and sensor
positions.
Â
Ü
Exit menu “Measure Place” menu
to adjust the Measurement Path Settings
.
1. Go to >Application< menu.
The following query appears on the display:
Fig. 23-3
Accept modified measurement place parameters