5-1
5
Maintenance Requirements
5.1
Tools Required
5.1.1 Tools/measurement devices and consumables
Table 5-1 Tools and measurement devices
Tools/measurement devices
Quantity
Remarks
Resin or stainless steel container
1
Large enough to accommodate two
transducers
Plastic bags
1
Approx. 2 m
×
2 m (0.3 mm in
thickness)
Vacuum cleaner
1
Blower (600W
~
16000 mm/min)
1
Makita 4014NV or equivalent
Cleaner
Brush
1
As small as a toothbrush
Air spray or blower brush
1
For cleaning the fans
Table 5-2 Consumables
Consumable
Quantity
Remarks
Aluminum foil
1
Physiological saline solution (0.85 to
0.95% saline solution)
-
The saline container should be
filled.
Glass cleaner
1
For cleaning the monitor
Neutral detergent
Soft rag
Summary of Contents for DC-6