System Structure and Assembly/Disassembly
4-51
c)
Replace the net and the net support with new ones.
4.3.21 Remove the transformer
a)
Remove the inlet dust net as per the step a in section 4.2.18.
b)
Remove 4 M3x8 screws of the separating plate of the transformer, and take out
the separating plate.
c)
Remove 4 inner hexagonal screws separating the transformer (2 screws at the
Net and support
4 M3x8 screws
Summary of Contents for DC-6
Page 1: ...DC 6 DC 6T DC 6Vet Diagnostic Ultrasound System Service Manual...
Page 2: ......
Page 20: ...2 1 2 System Overview 2 1 System Appearance 2 1 1 Complete System with CRT Monitor...
Page 23: ...System Overview 2 4 2 2 LCD Monitor...
Page 26: ...System Overview 2 7 2 2 3 Lever of upper support arm...
Page 66: ...4 1 4 System Structure and Assembly Disassembly 4 1 Exploded View of Complete System...
Page 101: ...System Structure and Assembly Disassembly 4 36 Power boards Card detacher...
Page 191: ...P N 2105 20 40473 V10 0...