Checks
6-2
version.
6.1.2.5
Check the system operation
Check that the menus and windows can pop up normally.
Check that measurements can be performed in each exam mode, such as OB measurements
and cardiac measurements, etc.
Check that exam modes can be normally switched.
Check that display modes can be normally switched.
Check that the transducers can be normally switched.
6.2
Function checks
6.2.1 Flowchart for function checks
Check the displayed time
and date
↓
Check the initial operations
↓
Check general operations
↓
Check keys on the control
panel
↓
Check the trackball
↓
Check the peripherals of the
main unit
↓
Check the fan
6.2.2 Content of the function checks
1)
Checking displayed time and date
Check that the displayed time and date are correct.
If not, correct them.