A-9
NIBP
Standards
Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1, EN1060-3, EN1060-4 and SP10
Technique Oscillometry
Mode of operation
Manual, Auto and STAT
Static pressure measurement
range
0kPa to 40.0kPa (0mmHg to 300mmHg)
Static pressure measurement
accuracy
±0.4kPa (±3mmHg)
Maximum measurement time
120s for adult and pediatric patients
90s for neonatal patients
Adult
Pediatric
Neonate
mmHg
40 to 270
40 to 200
40 to 135
Systolic
KPa
5.3 to 36.0
5.3 to 26.7
5.3 to 18.0
mmHg
10 to 210
10 to 150
10 to 100
Diastolic
KPa
1.3 to 28.0
1.3 to 20.0
1.3 to 13.3
mmHg
20 to 230
20 to 165
20 to 110
Measurement range
Mean
KPa
2.7 to 30.7
2.7 to 22.0
2.7 to 14.7
Setting range
Default
Accuracy
Adult:
80 to 280 mmHg
160 mmHg
Pediatric
80 to 210 mmHg
140 mmHg
Cuff inflation pressure
Neonate
60 to 140 mmHg
90 mmHg
5 mmHg
Adult:
297±3 mmHg
Pediatric: 240±3
mmHg
Software overpressure
protection
Neonate: 147±3
mmHg
Measurement accuracy
Max mean error:
±
5 mmHg
Max standard deviation: 8 mmHg
A.5
Power Supply Specifications
Fuse
Time-lag, 250V, T3.15A
AC power
Line voltage
100 to 240 VAC (±10%)
Current
1.8 to 0.8 A
Frequency
50 / 60Hz (±3Hz)
DC Power (with an external DC/AC adapter)
Input voltage
12VDC
Power consumption
190 W
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...