22-7
22.3.6
ECG Cable Test
It is recommended to perform ECG cable test once a year.
Test tool: ECG simulator
Follow this procedure to perform ECG cable test:
1.
Turn the Mode Select knob to “Monitor”.
2.
Connect the ECG cable to the defibrillator and the electrodes to the simulator.
3.
Turn on the simulator and select a normal ECG rhythm.
4.
Wait for a few seconds and check that the waveform is displayed normally and that no lead-off alarms reported in
the alarm information area.
22.3.7
Manual Defibrillation Test
Test tools: defibrillator/pacer analyzer
Charge/discharge
1.
Remove the batteries and connect the equipment with AC mains. Turn the Mode Select knob to Manual Defib.
2.
Connect the external paddles to the equipment and place the paddles on the defibrillator/pacer analyzer.
3.
Enter the Configuration Main screen. From the Record Setup menu set [
Shock Event
] to [
On
] so that shock events
can be recorded automatically if happened.
4.
Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as 0 or blank.
5.
Select the energy level to 1J.
6.
Charge/discharge the equipment to verify the energies measured by the analyzer meet the following accuracy:
Selected Energy (J)
Measured Value (J)
1
0 to 3
100
85 to 115
360
306 to 414
7.
Set the energy to 100J and 360J respectively. Repeat Step 6.
8.
Disconnect the equipment from the AC mains. Run the equipment on fully charged battery. Move the Mode Select
knob to Manual Defib. Repeat Steps 2 to 7.
9.
Verify that the equipment records the shock events automatically and correctly.
10. Use multifunctional electrode pads. Repeat Step 3 to Step 9.
Energy Disarming
1.
Run the equipment on fully charged battery. Move the Mode Select knob to Manual Defib.
2.
Connect the external paddles to the equipment and place the paddles on the defibrillator/pacer analyzer.
3.
Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as 0 or blank.
4.
Select the energy level to 360J.
5.
Charge the equipment.
6.
Verify that the charge tone is issued during charging.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...