22-8
7.
Press the “Disarm” soft key to discharge the energy internally.
8.
Verify that a prompt “Charge Removed” appears on the screen and the charge done tone stops.
9.
Verify that the value measured by the analyzer is 0J or blank.
10. Enter the Configuration Main menu, select [
Manual Therapy Setup
] and set [
Time to Auto Disarm
] to [
60s
].
11. Exit “Configuration Management”. The equipment restarts automatically.
12. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as 0 or blank.
13. Select the energy level to 360J.
14. Charge the equipment. Count time after charging is completed. Verify that the prompt “Shock Removed” appears
on the equipment and the energy measured by the analyzer is 0J or blank after 60 seconds.
15. Use multifunctional electrode pads. Repeat Steps 3 to 14.
Synchronous Defibrillation
1.
Connect the external paddles and ECG cable to the equipment. Place the paddles ECG electrodes on the
defibrillator/pacer analyzer.
2.
Set the analyzer to Time Measurement Mode and output normal sinus rhythms, e.g. amplitude value 1mV and HR
60bpm.
3.
Enter Configuration Management. In the [
Manual Therapy Setup
] menu, set [
Sync After Shock
] to [
On
].
4.
Adjust the energy setting of the equipment to be 10J.
5.
Press the [
Enter Sync]
soft key to start synchronous defibrillation. If Remote Sync is switched on, press the [
Enter
Sync]
soft key and select [
Local
] to start synchronous defibrillation
6.
Select Pads as the ECG source and begin charging.
7.
When charging finishes, press and hold the “Shock” button to deliver a shock.
8.
Verify that synchronous discharge succeeds and the delivered energy measured by the analyzer is 10J±2J.
9.
Verify that the delay time of synchronous defibrillation measured by the analyzer is less than 60ms.
10. Verify that the synchronous discharge mark appears on the R wave.
11. Verify that the prompt messages are correct during testing.
12. Select lead II as ECG source and perform charging. Repeat steps 7 to 11.
13. Use multifunctional electrode pads. Repeat steps 2 to 12.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...