19-8
19.3.10
Alarm Setup Menu
Menu Item
Options/Range
Default
Alarm Pause Time
1, 2, 3, 5, 10, 15 min, Permanent
2 min
Audio Off
Enabled, Disabled
Disabled
Alarm Volume
0 to 10 (If Audio Off is enabled),
1 to 10 (If Audio Off is disabled)
2
Reminder Tone
On, Off
Off
Reminder Volume
High, Med, Low
Med
High Level Alarm
3 to 15s
10s
Med Level Alarm
3 to 30s
20s
Alarm Tone
Interval
Low Level Alarm
15 to 100s
20s
Latching Alarms
Yes, No
No
Displaying Alarm Limits
Yes, No
Yes
ECG Lead Off Lev
High, Med, Low
Low
SpO2 Sensor Off Lev
High, Med, Low
Low
No Battery
Status Indicator On, Status Indicator Off
Status Indicator On
19.3.11
Waveform Setup Menu
Menu Item
Options
Default
Remark
Wave 1
/
ECG1
Unchangeable
Wave 2
Pleth, Resp
Pleth
Wave 3
Pleth, Resp
Resp
For optional parameters, options of
waveforms are available only when
the parameters are configured.
ECG Green
RESP Yellow
SpO
2
Cyan
Parameter/
Wave Color
NIBP
Green, Yellow, Cyan, White, Red, Blue, Purple,
Orange
White
/
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...