7-5
7.7
CPR
If [
Initial CPR Time
] is not configured as Off, the system enters initial CPR if AED mode is entered. You can set [
Initial
CPR Time
] to an appropriate time or switch it off through configuration management.
After the shock series, ECG analysis pauses and the equipment enters the CPR status. Analysis resumes at the completion
of the pause period or when you press the [
Resume Analyzing
] soft key in CPR status.
In current shock series, the equipment enters the CPR status if you press the [
Pause for CPR
] soft key after a shock is
delivered. The CPR pause time is defined by setting [
CPR Time
] through configuration management.
NOTE
z
You can start analyzing patient’s heart rhythm again at any time by pressing the [Resume Analyzing] soft
key in CPR status.
7.7.1
CPR Metronome
The equipment provides a CPR metronome feature that can be used to encourage rescuers to perform chest
compression and ventilation at AHA/ERC recommended rate.
When activated, the metronome sounds 120 times per minute and give voice prompts to indicate the rescuers to
perform ventilation at configured compression/ventilation rate.
To activate or deactivate CPR metronome,
1.
Move the Mode Select knob to Monitor, Manual Defib or Pacer. Press the Main Menu button on the front panel, and
then select
→
[
Others >>
]
→
[
Configuration >>
]
→
Enter the required password.
2. Select
[
AED Setup
] [
→
CPR Metronome
], and toggle between [
On
] and [
Off
].
The CPR metronome is activated by default. When CPR metronome is on, you can set the compression/ventilation rate
by selecting [
CPR Mode
]. The default rate is 30:2.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...