A-8
PR
Measurement range
20 to 254 bpm
Accuracy
±3 bpm (measured without motion)
±5 bpm (measured with motion)
Masimo SpO
2
Module
Measurement range
1 to 100%
Resolution 1%
Accuracy
70 to 100%:
70 to 100%:
70 to 100%:
1% to 69%:
±2% (measured without motion in adult/pediatric mode)
±3% (measured without motion in neonate mode)
±3% (measured with motion)
Not specified
Refreshing rate
1 s
SpO
2
averaging time
2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s
Low perfusion conditions
Pulse amplitude: >0.02%
Light penetration: >5%
Low perfusion SpO
2
accuracy
±2%
PR
Measurement range
25 to 240 bpm
Accuracy
±3 bpm (measured without motion)
±5 bpm (measured with motion)
Low perfusion PR accuracy
±3 bpm
Nellcor SpO
2
Module
Measurement range
0 to 100%
Resolution 1%
Sensor
Range
Accuracy*
MAX-A, MAX-AL, MAX-N
MAX-P, MAX-I, MAX-FAST
70 to 100%
0% to 69%
±2%
Not specified
OxiCliq A, OxiCliq N
OxiCliq P, OxiCliq I
70 to 100%
0% to 69%
±2.5%
Not specified
D-YS, DS-100A, OXI-A/N, OXI-P/I
70 to 100%
0% to 69%
±3%
Not specified
Accuracy
MAX-R, D-YSE, D-YSPD
70 to 100%
0% to 69%
±3.5%
Not specified
Refreshing rate
1 s
PR
Measurement range
20 to 300 bpm
Accuracy
±3 bpm (20 to 250 bpm)
Not specified (251 to 300 bpm)
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...