19-6
19.3.6
Resp Setup Menu
Menu Item
Options/Range
Default
Remark
RR Alarm
On, Off
Off
/
RR Alm Lev
High, Med, Low
Med
/
Sweep 6.25/12.5/25
mm/s
6.25mm/s
/
Adu 30
/
Ped
(Low+2) to 100
30 /
RR High
Neo
(Low+2) to 150
100
/
Adu 8
/
Ped 8
/
RR Low
Neo
6 to (High-2)
30 /
Apnea Time
10/15/20/25/30/35/40s
20s
/
19.3.7
SpO
2
Setup Menu
Menu Item
Options/Range
Default
Remark
SpO2 Alarm
On, Off
On
/
SpO2 Alm Lev
High, Med
Med
/
Sweep
12.5mm/s, 25 mm/s
25 mm/s
/
Adu 100
Ped 100
SpO2 High
Neo
(Low+1) to 100
95
/
Adu 90
Ped 90
SpO2 Low
Neo
Desat Limit to (High-1)
90
/
Adu 80
Ped 80
Desat
Neo
50 to (High-1)
80
/
Averaging Masimo
SpO
2
2-4s, 4-6s, 8s, 10s, 12s, 14s, 16s
8s
For Masimo SpO
2
module only.
Mindray SpO
2
High, Med, Low
Med
Sensitivity
Masimo SpO
2
Normal,
Maximum
Normal
Different options are available to
match the SpO
2
module used.
Pitch Tone
Mode 1, Mode 2
Mode 1
/
Sat-Seconds
0s, 10s, 25s, 50s, 100s
0s
For Nellcor SpO
2
module only.
19.3.8
PR Setup Menu
Menu Item
Options/Range
Default
Remark
PR Alarm
On, Off
Off
/
PR Alm Lev
High, Med, Low
Med
/
Adu 120
Ped 160
PR High
Neo
(Low+2) to 240
200
PR ranges are different for different
modules.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...