Remote Synchronized Cardioversion
The equipment can be configured to receive an ECG source from a remote patient monitor (such as a bedside patient
monitor) to perform synchronized cardioversion. To do so, the remote patient monitor shall have a sync out connector
and shall be connected to the defibrillator/monitor’s multifunctional connector with a synchronous cable.
To switch on remote synchronization, select [
Manual Defib Setup
]
→
[
Remote Sync
]
→
[
On
] through the Configuration
Main Menu.
The remote synchronization procedure is as follows:
1.
Use a sync cable to connect the defibrillator/monitor with a bedside monitor through the defibrillator/monitor’s
multifunctional connector.
2.
Switch on the equipment and enter the Manual Defib mode.
3.
Press the [
Enter Sync
] soft key. In this case the “
Select Sync Mode
” dialog pops up.
4. Select
[
Remote
] to enter remote synchronization. Then the message “
Remote Sync
” is presented.
5.
Confirm a square wave on the defibrillator/monitor blinks with each detected R wave on the remote monitor as
shown below, indicating a sync pulse is received.
6.
Connect the therapy cable to the therapy port. Push until you hear it click into place.
7.
Apply multifunction electrode pads or external paddles to the patient.
8.
Follow Steps 6 through 9 from
8.5.1 Performing Synchronized Cardioversion
as described previously.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...