22-11
22.3.9.4
NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at least once every two years or
whenever you doubt the NIBP reading.
Tools required:
An adult cuff
An air tubing
A correct sized cylinder
Follow this procedure to perform the leakage test:
1.
Set the patient category to [
Adu
].
2.
Connect the cuff to the equipment’s NIBP connector.
3.
Wrap the cuff around the cylinder as shown below.
4.
In the Installation Mode Main menu, select [
Maintain
NIBP
]
→
[
Start Leakage Test
].
After about 20 seconds, the equipment automatically deflates. This means the leakage test is started.
When the accuracy test is completed, the result will be displayed. If the message “NIBP Pneumatic Leak” is displayed, it
indicates that the NIBP airway may have leakages. Check the tubing and connections for leakages, and then perform a
leakage test again.
If the problem persists, contact your service personnel.
When you select the [
Start
Leakage Test
] button, it turns to be [
Stop
Leakage Test
]. Select [
Stop
Leakage Test
],
leakage test stops and the button turns to be [
Start
Leakage Test
] again.
22.3.9.5
Checking Version Information
To view version information, select [
Version
] in the Installation Mode Main menu. In the popup menu, you can view
system software version and module software version.
Defibrillator/monitor
Connector for NIBP
cuff
Tubing
Cylinder
Cuff
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...