B-2
Guidance and Declaration - Electromagnetic Immunity
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
(EFT)
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
(length greater than 3 m)
±2 kV for power supply lines
±1 kV for input/output lines
(length greater than 3 m)
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% UT (>95% dip in UT) for
0.5 cycle
40% UT (60% dip in UT) for 5
cycle
70% UT (30% dip in UT) for
25 cycle
<5% UT (>95% dip in UT) for
5 seconds
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT) for
5 cycle
70% UT (30% dip in UT) for
25 cycle
<5% UT (>95% dip in UT)
for 5 seconds
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of our
product requires continued
operation during power mains
interruptions, it is recommended
that our product be powered from
an uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
Note: UT is the A.C. mains voltage prior to application of the test level.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...