20-2
20.2
Installing the Batteries
To install the batteries:
1.
Align a battery with the battery compartment.
2.
Insert the battery, and press until you hear it click into the place.
To replace a battery, press the latch on the battery and push the battery to the right until you remove it. Insert a new
battery into the battery compartment.
20.3
Battery Alarms
20.3.1
No Battery Alarm
You can configure the status of status indicator when no battery is installed in the configuration mode.
Press the Main Menu button on the front panel of the equipment. Select [
Others >>
]
→
[
Configuration >>
]
enter the
→
required password. In the Configuration main menu, select [
Alarm Setup
]
→
[
No Battery
], and toggle between [
Status
Indicator On
] and [
Status Indicator Off
]. The default setting is [
Status Indicator Off
].
If [
Status Indicator On
] is selected, the red cross status indicator flashes in the case that battery is not installed,
meanwhile a message “
No Battery
” is presented in the Technical Alarm Area.
If [
Status Indicator Off
] is selected, the red cross status indicator will not be illuminate in the case that battery
is not installed; the message “
No Battery
” will be presented in the Technical Alarm Area.
20.3.2
Low Battery Alarm
If the equipment is run on battery and the battery charge is low, a technical alarm “
Low Battery
” will be triggered. In this
case, replace the battery with a fully charged on or connect the equipment to external power supply immediately.
If the battery is almost depleted, a prompt “
Battery depleted! System will shut down imminently. Connect to AC
mains or replace battery
” pops up. In monitoring, manual defibrillation, and pacing mode, additional alarm lights and
alarm tones are provided. In this case, take appropriate actions immediately. This prompt will not disappear until the
battery is replaced or the equipment is connected to the external power supply. The equipment automatically shuts
down if no action is taken within a period of about 3 minutes.
NOTE
z
The Low Battery alarm means that the battery is beginning to weaken and should be replaced at the first
opportunity. At least 20 minutes of monitoring and six full energy shocks can be performed when the Low
Battery alarm is activated. Replace the battery or connect the equipment to AC mains as soon as possible.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...