19-5
Menu Item
Options/Range
Default
Remark
Arrhythmia On,
Off
Off
/
ARR Alarm
On, Off
On
/
PVCs/min
R ON T
VT>2
Couplet
PVC
Vent. Rhythm
Bigeminy
Trigeminy
Tachy
Brady
PNP
PNC
Missed Beat
Multif. PVCs
Nonsus. Vtac
Pause
ARR Alm Lev
Irr. Rhythim
High, Med, Low
Med
/
Asystole Delay
3 to 10
5
/
V-Tach Rate
100 to 200
130
/
V-Tach PVCs
3 to 99
6
/
Vbrd Rate
15 to 60
40
/
Vbrd PVCs
3 to 99
5
/
Adu
60 to 300
160
/
Extreme
Tachy
Ped
60 to 300
180
/
Adu
15 to 120
35
/
Extreme
Brady
Ped
15 to 120
50
/
PVCs High
1 to 10
10
/
Adu
60 to 300
120
/
Tachy
Ped
60 to 300
160
/
Adu
15 to 120
50
/
Brady
Ped
15 to 120
75
/
Multif. PVCs Window
3 to 31
15
/
Pause Time
1.5, 2.0, 2.5
2.0
/