13-3
Warning
z
Continuous CUFF pressure due to connection tubing kinking may cause blood flow interference and
resulting harmful injury to the patient.
13.5.2
Starting and Stopping NIBP Measurements
You can start and stop NIBP measurements by using the
hardkey on the equipment’s front panel.
13.5.3
Correcting the Measurement
The cuffed limb should be at the same level as the patient’s heart. If not, correct the measurement by:
Add 0.75 mmHg (0.10 kPa) for each centimeter higher, or
Deduct 0.75 mmHg (0.10 kPa) for each centimeter lower.
13.5.4
Enabling NIBP Auto Cycling
1.
Select the NIBP parameter area to access the [
NIBP Setup
] menu.
2. Select
[
Interval
] and then select a desired time interval. Selecting [
Manual
]
switches to manual mode.
3.
Start the measurement manually. The equipment will then automatically repeat NIBP measurements at the set time
interval.
Warning
z
Continuous non-invasive blood pressure measurements may cause purpura, ischemia and neuropathy in
the limb with the cuff. Inspect the application site regularly to ensure skin quality and inspect the extremity
of the cuffed limb for normal color, warmth and sensitivity. If any abnormity occurs, move the cuff to
another site or stop the blood pressure measurements immediately.
13.5.5
Starting a STAT Measurement
1.
Select the NIBP parameter area to access the [
NIBP Setup
] menu.
2. Select
[
NIBP STAT
].
The STAT mode initiates a 5-minute continuous, automatic NIBP measurement.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...