D-3
Measurement
Alarm Message
L
I
Cause and solution
SpO2 Too Much Light
L
C
There is too much light on the SpO
2
sensor. Move the sensor
to a place with lower level of ambient light or cover the
sensor to minimize the ambient light.
SpO2 Low Signal
L
C
SpO2 Weak Signal
L
C
SpO2 Weak Pulse
L
C
SpO2 Low Perf
L
B
The SpO
2
signal is too low or too weak. Check the patient’s
condition and change the sensor application site. If the error
persists, replace the sensor.
SpO2 Interference
L
C
The SpO
2
signal has been interfered. Check for any possible
sources of signal noise form the area around the sensor, and
check the patient for excessive motion.
SpO2 Non-Pulsatile
L
C
SpO2 Board Fault
L
C
There is a problem with the SpO
2
measurement board. Do not
use the module and contact your service personnel.
NIBP Loose Cuff
L
A
NIBP Air Leak
L
A
The NIBP cuff is not properly connected, or there is a leak in
the airway.
NIBP Pneumatic Leak
L
A
Check the NIBP cuff and pump for leakages.
NIBP Cuff Type Wrong
L
A
The cuff type applied mismatches the patient category. Verify
the patient category and replace the cuff.
NIBP Air Press. Err
L
A
An error occurred to the air pressure. Verify that the
equipment application site meets the environmental
requirements and check if there is any source that affects the
air pressure.
NIBP Weak Signal
L
A
The patient’s pulse is weak or the cuff is loose. Check the
patient’s condition and change the cuff application site. If the
problem persists, change the cuff.
NIBP Sig. Saturated
L
A
The NIBP signal is saturated due to excess motion or other
sources.
NIBP Overrange
L
A
The patient’s NIBP value may be beyond the specified
measurement range.
NIBP Excessive Motion
L
A
Check the patient’s condition and reduce the patient motion.
NIBP Cuff Overpress
L
A
The NIBP airway may be occluded. Check the airway and
measure again.
NIBP Equip Err
H
A
NIBP Time Out
L
A
NIBP Measure Failed
L
A
An error occurred during NIBP measurement and therefore
the equipment cannot perform analysis correctly. Check the
patient’s condition and NIBP connections, or replace the cuff.
NIBP
NIBP Reset For Err
L
A
An illegal reset occurred during NIBP measurement. Check if
the airway is occluded.
No Fan
L
C
Make sure that the fan is connected.
No Speaker
L
C
Make sure that the speaker is connected.
No Data Card
L
C
Make sure that the CF card is in place or format the CF card.
Main control
system
Power Board Comm Err
H
C
An error occurred to the power board, or there is a problem
with the communications between the power board and the
host. Restart the equipment.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...