B-3
Guidance and Declaration - Electromagnetic Immunity
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment - guidance
3 Vrms
150k to 80 MHz
Outside ISM
bandsa
3 Vrms (V1)
Conduced RF
IEC 61000-4-6
10 Vrms
150kHz to
80MHz
in ISM bandsa
(for life support
devices)
10 Vrms (V2)
10V/m
80 MHz to 2.5
GHz
(for life support
devices)
10 V/m (E1)
Radiated RF
IEC 61000-4-3
20V/m
80 MHz to 2.5
GHz
(ISO 9919)
20 V/m (E2)
Portable and mobile RF communications equipment should
be used no closer to any part of the device, including cables,
than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance:
P
V
d
⎥⎦
⎤
⎢⎣
⎡
=
1
5
.
3
P
V
d
⎥⎦
⎤
⎢⎣
⎡
=
2
12
P
E
d
⎥⎦
⎤
⎢⎣
⎡
=
1
12
80 MHz to 800 MHz
P
E
d
⎥⎦
⎤
⎢⎣
⎡
=
1
23
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in meters (m)b.
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site surveyc, should be less than the
compliance level in each frequency range d
Interference may occur in the vicinity of equipment marked
with the following symbol:
.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...