7-2
WARNING
z
During defibrillation, air pockets between the skin and multifunction electrode pads can cause patient skin
burns. To help prevent air pockets, make sure defibrillation pads are completely adhered to the skin.
z
Do not use dried-out pads.
z
In AED mode, this equipment is not designed to administer energy at pediatric joule settings. The American
Heart Association recommends AEDs be used only on patients who are more than eight years old.
CAUTION
z
Aggressive handling of multifunction electrode pads in storage or prior to use can damage the pads. Discard
the pads if they become damaged.
z
For patients with implantable pacemaker, the sensitivity and specificity of AED algorithm may be impaired.
7.3
AED View
A typical screen in AED Mode is shown below.
In AED mode, HR numeric and one ECG waveform acquired from the multifunction electrode pads are displayed. Below
the ECG is the information area which displays the defibrillation mode, prompt message, contact impedance indicator,,
selected energy and a shock counter.
For details about the contact impedance indicator, refer to
8.7 Contact Impedance Indicator
.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...