E-2
E.3
Device Labelling
Check the labels provided by the manufacturer or the healthcare facilities are present and legible.
Main unit label
Integrated warning labels
E.4
Protective Earth Resistance
1.
Plug the probes of the analyzer into the device’s protective earth terminal and protective earth terminal of the AC
power cord.
2.
Test the earth resistance with a current of 25 A.
3.
Verify the resistance is less than limits.
LIMITS
For all countries, R = 0.2 Ω Maximum
E.5
Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other leakage tests.
The following outlet conditions apply when performing the Earth Leakage test:
normal polarity (Normal Condition)
reverse polarity (Normal Condition)
normal polarity with open neutral (Single Fault Condition)
reverse polarity with open neutral (Single Fault Condition)
LIMITS
For UL60601-1,
300 μA in Normal Condition
1000 μA in Single Fault Condition
For IEC60601-1,
500 μA in Normal Condition
1000 μA in Single Fault Condition