19-7
Menu Item
Options/Range
Default
Remark
Adu 50
Ped 75
PR Low
Neo
25 to (High-2)
100
QRS Volume
0 to 10
2
This setting is linked with the [
QRS
Volume
] setting in the [
ECG Setup
]
menu
19.3.9
NIBP Setup Menu
Menu Item
Options/Range
Default
Remark
Interval
Manual, 1 min, 2 min, 2.5 min, 3 min, 5
min, 10 min, 15 min, 20 min, 30 min, 1 h,
1.5 h, 2 h, 3 h, 4 h, 8 h
Manual
/
Press. Unit
mmHg, kPa
mmHg
Alarm limits are refreshed in real time
when pressure unit is changed.
NIBP-Sys
NIBP-Dia
Alarm
NIBP-Mean
On, Off
On
/
Alm Lev
High, Med
Med
/
Adu
(Low + 5) to 270
160
Ped
(Low + 5) to 200
120
Sys High
Neo
(Low + 5) to 135
90
/
Adu 90
/
Ped 70
/
Sys Low
Neo
40 to (High - 5)
40 /
Adu
(Low + 5) to 230
110
/
Ped
(Low + 5) to 165
90
/
Mean High
Neo
(Low + 5) to 110
70
/
Adu 60
/
Ped 50
/
Mean Low
Neo
20 to (High - 5)
25 /
Adu
(Low + 5) to 210
90
/
Ped
(Low + 5) to 150
70
/
Dia High
Neo
(Low + 5) to 100
60
/
Adu 50
/
Ped 40
/
Dia Low
Neo
10 to (High - 5)
20 /
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...