9-2
CAUTION
z
Use of Pacer mode may be password protected. Make sure the operator knows and remembers the
password as defined in Configuration. Failure to enter correct password will prevent the delivery of pacing
therapy.
z
For treatment of patients with implanted devices such as permanent pacemakers or
cardioverter-defibrillators, consult a physician and the instructions for use provided by the device’s
manufacturer
z
Prolonged noninvasive pacing may cause patient skin irritation and burns. Periodically inspect the
underlying skin and change ECG electrodes and multifunction electrode pads.
NOTE
z
If pacing is interrupted for any reason, the [Start Pacing] soft key must be pressed to resume pacing.
z
In pacer mode, you cannot change the patient’s internal paced status from the ECG Setup menu.
z
In the case that pads poorly contact the patient, the alarm “Pacer Stopped Abnormally” and “Pads Off” may
be presented.
z
Pads are not an available choice for the source of ECG waveform in Pacer mode.
9.3
Pacing View
A typical screen in pacer Mode is shown below.
In pacer mode, an ECG waveform, related parameters and waveforms are displayed. The pacer information area shows
the pacer mode, pacer rate and output, as well as pacer-related prompts and alarms. Soft keys available for setting
pacing functions are also displayed.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...