8-6
NOTE
z
When internal paddles are used for defibrillation, the energy selection is automatically limited to 50 joules
because of possible cardiac damage from higher energies.
z
Sterilize the internal paddles before each use. Otherwise, severe infection may result.
z
Clean the internal paddles after each use.
8.5
Synchronized Cardioversion
Synchronized Cardioversion allows you to synchronize delivery of the defibrillator shock with the R-wave of the ECG. You
may choose to perform synchronized cardioversion through either:
Multifunction electrode pads, or
External paddles
To use synchronized cardioversion, press the [
Enter Sync
] soft key in the asynchronous defibrillation mode. Then “Sync”
appears in the manual Defibrillation information area and a marker appears above each R-wave, see the figure below:
You can monitor ECG through multifunction electric pads, or external paddles, or electrodes connected to a 3- or 5-lead
ECG cable. Shock is delivered through either pads or paddles.For synchronized cardioversion, we recommend to acquire
patient’s ECG through ECG lead set
NOTE
z
When you enter synchronous cardioversion, monitoring alarms is reactivated autonomously.
Sync marker
R-wave marker
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...