5-3
5.3.2
Audible Alarms
The equipment uses different alarm tone patterns to match the alarm level:
High level alarms
double beeps.
Medium level alarms
triple beeps.
Low level alarms
single beep.
5.3.3
Alarm Message
When an alarm occurs, the alarm message will appear in the technical or physiological alarm area. For physiological
alarms, the asterisk symbols (*) before the alarm message match the alarm level as follows:
High level alarms
***
Medium level alarms
**
Low level alarms
*
Additionally, the alarm message has different background color which matches the alarm level.
For physiological alarms
High level alarms
red
Medium level alarms
yellow
Low level alarms
yellow
For technical alarms
High level alarms:
red
Medium level alarms:
yellow
Low level alarms:
blue
5.3.4
Flashing Numeric
If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in alarm will flash every second,
and corresponding alarm limit will also flash at the same frequency indicating the alarm limit is violated.
5.3.5
Alarm Status Symbols
Apart from the aforementioned alarm indicators, the equipment still uses the following symbols telling the alarm status:
indicates all alarms are paused.
indicates all system alarms are turned off.
indicates alarm sounds are paused.
indicates alarm sounds are turned off.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...