D-4
Measurement
Alarm Message
L
I
Cause and solution
Keyboard Comm Err
L
C
An error occurred to the keypad board, or there is a problem
with the communications between the keypad board and the
host. Restart the equipment.
Therapy Module Comm Err
S
C
An error occurred to the therapy module, or there is a
problem with the communications between the therapy
module and the host. Restart the equipment. If the problem
persists, contact your service personnel.
Main Control Selftest Err
H
C
The main control voltage is abnormal. Contact your service
personnel.
RT Clock Need Reset
L
C
Reset system time.
RT Clock Err
H
C
An error occurred to the RTC chip, or the button cell is
depleted. Replace corresponding part.
Data Card Err
L
C
There is a problem with the data card. Check or replace the
data card if necessary.
Power System Selftest Err
H
C
Power Board Volt Err
L
C
An error occurred to the system power supply. Restart the
equipment.
No battery
L
C
No battery is installed. Please install the battery.
Low Battery
S
C
Change battery or connect the equipment to the AC power
source to charge the batteries.
Battery Err
H
There is a problem with the batteries. Check the batteries for
damage; verify that correct batteries are used. Replace the
batteries if necessary.
Battery Depleted! System
will shut shown imminently.
Connect to AC Mains or
Replace Battery.
S
C
Connect the equipment to AC mains.
Battery Aged
L
C
Replace the battery.
Power board
Battery failed charging
M
C
Battery failure or power board hardware failure. Replace the
battery. If the problem persists, contact your service
personnel.
Therapy Equip selftest Err
S
C
An error occurred during therapy module self test. Restart the
equipment If the problem persists, contact your service
personnel.
Defib Malfunction
S
C
The defibrillation function fails or both the defibrillation and
pacing functions fail. Restart the equipment and test
defibrillation function. If the problem persists, contact your
service personnel.
Pacer Malfunction!
S
C
The pacing function fails. Restart the equipment and test
pacer function. If the problem persists, contact your service
personnel.
Disarming Failed
H
C
There is a problem with the therapy module disarming circuit.
Contact your service personnel.
Last User Test Failed
L
C
Run a successful user test.
Therapy
module
Last Auto Test Failed
L
C
Run a successful user test again.
Summary of Contents for BeneHeart D3
Page 1: ...BeneHeart D3 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 16: ...8 FOR YOUR NOTES...
Page 22: ...1 6 FOR YOUR NOTES...
Page 40: ...4 2 FOR YOUR NOTES...
Page 62: ...6 12 FOR YOUR NOTES...
Page 78: ...8 10 FOR YOUR NOTES...
Page 88: ...10 4 FOR YOUR NOTES...
Page 90: ...11 2 FOR YOUR NOTES...
Page 102: ...14 2 FOR YOUR NOTES...
Page 170: ...D 6 FOR YOUR NOTES...
Page 182: ...G 2 FOR YOUR NOTES...
Page 184: ......
Page 185: ......
Page 186: ...P N 046 001653 00 6 0...