SERVICE MANUAL
4-12
9G9002 – 2011-08-15
Condition Code: 01-005 – Error Stopping a Subsystem
Condition Code: 01-007 – General Subsystem Communication Error
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
None
Error stopping
subsystem xxxx (where
xxxx is the subsystem
name)
NORMAL
1-Alert
Description
An error occurred while attempting to stop a subsystem.
System Impact
The subsystem in question may still be running, in which case full shutdown will
not be possible.
User Action
None
Show at DS
False
DICOM Printer
Status
NORMAL
Unique Symb. ID
ERR_SUBSYSTEM_STOP
Note
Rationale for not reporting to user: The user requests shutdown of the entire
laser imager, not just a subsystem. The OS will eventually shut down regardless
of whether all subsystems shut down successfully.
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
None
Configuration and
Control Subsystem
cannot communicate
with subsystem xxxx
(where xxxx is the
subsystem name)
NORMAL
2-Warning
Description
An unknown error occurred while communicating with a subsystem.
System Impact
Configuration-and-Control will attempt to re-establish communications with the
subsystem in question. If communication is lost, the subsystem will be shut
down by force and restarted.
User Action
None
Show at DS
False
DICOM Printer
Status
NORMAL
Unique Symb. ID
ERR_SUBSYSTEM_COM