SERVICE MANUAL
3-8
9G9002 – 2011-08-15
Resetting the PM Counts and Changing the Service History
[1]
In the service tool, select
Monitoring and Control > Imager Monitor
.
The Imager Monitor screen displays.
[2]
Click:
•
Reset Prints To PM
This action also enters the PM record into the service history.
•
Reset Prints To Filter Replacement
[3]
Perform
“Completing the PM”
.
Completing the PM
[1]
Wait while the processor warms to Ready.
[2]
Check the temperature of the drum (see Adjustments and Replacements,
“Processor—Drum Temperature
Adjustment”
).
[3]
If necessary, adjust the temperature.
[4]
Make a flat-field gray test print.
[5]
Compare the test print with the test print made before the PM.
[6]
Make a print of a customer image.
[7]
Check the quality of the print with the customer.
[8]
Leave 2 deodorant filters with the customer.