Diagnostics
9G9002 – 2011-08-15
4-29
Condition Code: 06-420 – Invalid Data or Data Access Error
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
MIM Core: Internal
Software Error
• Page width and/or height are
less than zero
• Invalid page band
• Invalid coordinates for
indicator icon: (X1,Y2),
(X2,Y2). (where X1, Y1 and
X2, Y2 are the calculated
icon coordinates)
• Image size mismatch error
• Error in cropping coordinates
• The file XXXX could not be
found (where XXXX is the file
name)
• Error reading image XXXX
(where XXXX is the name of
the image file)
• General error accessing
database components
• General error accessing
page formatter
NORMAL
1-Alert
Description
The page builder component has encountered the following data error:
• The specified page width and/or height is less than zero.
• The number of page bands is neither 1 nor 3, and each of the bands does
not have the same type.
• Icon (x1,y1) and/or (x2,y2) is beyond the page extents.
• Image size mismatch error, where the size read does not match the size
expected.
• Bad cropping coordinates.
• Icon or image file not found.
• Trying to read image data; most likely, the image input file could not be
found.
• General error accessing the database component.
System Impact
The current page will not be processed because it could not be rendered due to
an internal error.
User Action
1. Resend the print job from the image source.
2. If the error persists, call for service.
Show at DS
True
DICOM Printer
Status
NORMAL
Unique Symb. ID
ERR_INVALID_DATA
Note
The log entry reflects the specific cause from the listed set of causes.
Current page will not be delivered. (Undeliverable)