Diagnostics
9G9002 – 2011-08-15
4-57
Condition Code: 21-632 – Dmax Not Met
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
Film Supply: Film
Calibration Failure -
Dmax Outside Target
Film Supply:
Calibration failed -
Dmax Not Met
CALIBRATION ERR
1-Alert
Description
The error occurs when the calibration sheet does not have a step density at or
above the required Dmax value. The most likely causes are expired film or
incorrect processor temperature. The error is cleared when a new calibration is
initiated or when the film cartridge is removed. Any cartridge using these
calibration results will automatically go into AIQC-Manual Mode and will
continue printing but with reduced range in achievable densities. A new
calibration will be attempted when this cartridge is re-inserted.
System Impact
None. The system will function normally. [SI-0]
User Action
The maximum density of the film is too low. Calibration results for this film
are outside the normal range but will still be used.
If you do not wish to use these calibration results, perform a new calibration
on this film or try another film cartridge.
FE Action
Use the service tool to check the processor temperature and set point for
correct values. If necessary, adjust the temperature (see Adjustments and
Replacements,
“Processor—Drum Temperature Adjustment”
).
Show at DS
True
DICOM Printer
Status
WARNING
Unique Symb. ID
ERR_UP_DMAX_NOT_MET
Logged for RMS
[CartridgeInfo=bbeeee-cjj-d-iii][FilmSize=size][TriggerValues=a,b,c,...,u]
[TriggerNotes=Optical Density values, from max to min]