SERVICE MANUAL
1-12
9G9002 – 2011-08-15
[10]
Will you set the clock manually, or does the site use a time server for synchronization?
[11]
Click
. The System Identification screen appears.
Note
Completing this screen is optional.
[12]
Enter the following information:
• Country Code
• Postal Code
• Hospital Name
• Department Name
• Device Location
[13]
Click
Save
.
[14]
Click
to return to the main screen.
[15]
Select
Utilities > Session > Restart
, and then click
Restart
.
Note
Make sure to restart using these steps or the network parameters will not be saved.
Wait for the laser imager to restart. The laser imager requires approximately 30 minutes to warm up. After completing
the warm-up cycle, the laser imager may print a calibration film if the film cartridge requires it. While the las
er imager
is warming up, continue with the next step.
Manually
Synchronization
a. Select the time zone.
b. Edit the time if needed.
c. Click
Save
.
a. Select the
Enable Time
Synchronization
check box.
b. In the IP Address field, type the
address of the time server to be used
for synchronization.
c. In the Period field, select the
synchronization period (how often the
clock is synchronized) from the drop-
down menu.
d. Select the time zone.
e. Click
Save
.