Preventive Maintenance
9G9002 – 2011-08-15
3-7
Caution
Isopropyl alcohol is highly flammable. Follow the manufacturer’s instructions for safe use and handling.
[19]
Use alcohol wipes and a scraper to inspect and clean debris and buildup from the:
• Heat shield, including the bottom side of the shield
• Cooling plate
• Densitometer guide
[20]
Vacuum the:
• Top cover
• Film supply
• Entire processor assembly
[21]
Install a new felt pad assembly with 2 screws (see Adjustments and Replacements,
“Felt Pad Assembly”
).
[22]
Carefully remove the film you inserted over the imaging roller and discard it.
[23]
Close the heat shield.
[24]
Install the 2 rear screws on the processor drum cover.
[25]
Install the:
• New deodorant filter
• Right cover
• Left cover
• Top cover
[26]
Close the film supply.
[27]
Place all used alcohol wipes and cleaning pads in a bag, close the bag with a tie-wrap, and then dispose of it
according to local ordinances.
[28]
Energize the imager.
[29]
Perform
“Resetting the PM Counts and Changing the Service History”
.
Heat Shield
Cooling Plate
Densitometer
Guide