Diagnostics
9G9002 – 2011-08-15
4-73
Condition Code: 29-925 – Densitometer Reference Reading Failed Diagnostics
Condition Code: 29-945 – Densitometer Reference Reading Diagnostics Warning
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
Densitometer: Internal
Diagnostic Failure
Densitometer reference
reading out of range
with light source on
ELEC DOWN
0-Fatal
Description
During a calibration print and during densitometer diagnostics, a reading from
the density sensor is taken with the light source turned on. This error occurs
when the reading is out of range.
Note
Code 29-945 is generated when this reading is nearing the range limit.
The error clears only when imager power is cycled.
System Impact
The imager can only print jobs that were in process when the error occurred.
The spooler can still accept jobs, but the jobs will not print until the error is
resolved. All jobs that were in process when the error occurred will reprint
completely when the error is resolved. [SI-1]
User Action
1. Shut down the laser imager, and then restart the imager.
2. If the error persists, call for service.
Show at DS
True
DICOM Printer
Status
FAILURE
Unique Symb. ID
ERR_DENSITOMETER_REFERENCE_LIGHT_ON
Logged for RMS
[TriggerValues=a, b, c, d][TriggerNotes=Density Reference, Density Offset,
Fiducial Reference, Fiducial Offset]
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
None
Densitometer reference
reading outside of
optimal range with light
source on
NORMAL
3-Notify
Description
During a calibration print and during densitometer diagnostics, a reading from
the density sensor is taken with the light source turned on. This notification
occurs when the reading is outside of the optimal range (approaching the range
limit).
Note
Code
29-925
is generated when this reading is over the range limit.
System Impact
None. The system will function normally. [SI-0]
User Action
None
Show at DS
False
DICOM Printer
Status
NORMAL
Unique Symb. ID
ERR_DENSITOMETER_OFFSET_LIGHT_ON_NOTICE