SERVICE MANUAL
2-4
9G9002 – 2011-08-15
To Adjust:
[1]
Log on to the service tool (see Additional Service Procedures,
“Logging On to the Service Tool”
).
[2]
Select
Configuration > Hardware > Processor Configuration
.
[3]
Click
Modify
.
[4]
In the Temperature field, enter the temperature measured by the meter.
Note
The Offset field indicates the offset value between the Set Point field and the Temperature field.
[5]
Click
Set Temperature
.
[6]
Allow approximately 10 minutes for the temperature of the processor drum to become stable.
[7]
Use the meter to check the temperature of the processor drum. The temperature must be 123.5–124.5 °C.
[8]
When the temperature is between 123.5–124.5 °C, close the Processor Configuration window.
[9]
Log off from the service tool (see Additional Service Procedures,
“Logging Off the Service Tool”
).
[10]
Install the back cover.